Once start is hit, open up the Triggered Run window (Window 1). Click Connect. Once
the system is connected, click Start. The TGA/MS run is now complete and should run.
If the first experiment in the MS did not run, there was too long of a delay
between hitting Start in the Triggered Run window.
If the subsequent experiments did not run, there was too long of a delay
between the end of the first experiment and the start of the second
experiment. If the method 25-500 C; 10C/m TGA/MS auto; 70uL alumina
was used, this should not occur. If this method was used and this occurred,
contact the person in charge of the instrument to determine if a solution
can be found.